FAQ: What are the operating and storage temperature limits for the N3300 electronic load family?

Storage Temp Max = 70°C; (Environmental Lab tests to 70°C) instrument is spec’d to operate (de-rated) at 55°C ambient.

Storage Temp Min = -40°C; (Environmental Lab tests to -40°C), caps and LCD display are only potential concerns going to -40°C]

Applies to: N3300A, N3301A, N3302A, N3303A, N3304A, N3305A, N3306A, N3307A

Select Home to return to the table of contents for more FAQs.

Visit our Power Products Forum for additional topics and discussions.

Read technical postings on our Power Blog – Watt’s Up?